Mini Maxwell Support
Known Issues

The following is a list of known issues.  Please send email to This e-mail address is being protected from spambots. You need JavaScript enabled to view it This e-mail address is being protected from spam bots, you need JavaScript enabled to view it for more information or to report further issues.

We also suggest that you take a look at FAQ.

Software updates for Mini Maxwell are found in the Customer Downloads pages.

  1. Disappearing Impairments

    Date posted: March 21, 2008

    Comments updated: March 27, 2008

    Revision containing correction: 3

    Summary: It has been noticed that sometimes the actual impairments return to the no-impairment state when moving from the web page to web page.

    Status: The problem has been isolated; a fix has been implemented and incorporated into the software update package.

  2. Static IP Configuration Failure

    Date posted: March 25, 2008

    Comments updated: March 27, 2008

    Revision containing correction: 3

    Summary: Static IP configuration does not work.

    Status: The problem has been isolated; a fix has been implemented and incorporated into the software update package.

  3. Delay Timing - 5 millisecond bias

    Date posted: March 27, 2008

    Comments updated: April 9, 2008

    Revision containing correction: 4.

    Summary: 5 Millisecond difference in delay between what user requests and what actually occurs.

    Details: When a user requests a delay of N milliseconds and N > 6 the actual delay is approximately N+5 milliseconds.  When N is zero the actual delay is zero.  When 0 < N < 6 then the actual delay is is approximately 5 to 6 milliseconds.

    Status: Resolved. Except as noted in the next sentence, delay is now rendered so that the average delay is within one millisecond of what the user requests.  When a user requests a delay of 1 or 2 milliseconds that will be rendered as approximately 3.5 milliseconds.

  4. Update Failure

    Date posted: December 31, 2008

    Comments updated: January 28, 2009

    Revision containing correction: 8.

    Summary: Update system fails due to failure to recognize USB hardware.

    Details: When doing a firmware update the update fails because the USB hardware is not recognized.  The update fails, the system is left unchanged.

    This problem exists only in Revision 7.

    This can not be repaired by the user.

    However, if the user is running Revision 7 there is no reason to do an update.

    Status: Resolved in Revision 8.

 